The Fluke Model 7250 pressure calibrator / controller (formerly Ruska 7250) offers an economical approach to automatic pressure testing and calibration with a precision of 0.003 % of full scale for ranges to 2,500 psi (17.2 MPa). A 3,000 psi (21 MPa) full scale range is available with 0.012% of reading precision from 900 psi to 3000 psi (6.3 to 21 MPa) precision.Models 7250xi and 7250i offer advanced percent of reading precision for increased capability with a single instrument, reducing the investment required to calibrate a wide variety of pressure devices and ranges.
The Model 7250xi provides 0.005 % of reading precision from 5 % to 100 % of the instrument’s range. This unmatched precision is achieved through unique quartz pressure sensing technology and multiple quartz sensors in a single instrument. Various ranges from 0 to 20 and 0 to 2 500 psi (0 to 140 kPa and 0 to 17.2 MPa) are available.
The Model 7250i provides 0.005 % of reading from 25 % to 100 % of range and is available in various ranges from 0 to 5 and 0 to 2,500 psi (0 to 35 kPa and 0 to 17.2 MPa).
For pressure below the lower threshold of 5% for the 7250xi and 25 % for the 7250i, the precision becomes 0.005 % of the lower value. For example, a 1,000 psi (7 MPa) Model 7250xi provides 0.005 % of reading from 50 to 1,000 psi (0.035 to 7 MPa); the precision for pressures from 0 to 50 psi (0 to 0.035 MPa) is 0.005 % of 50 psi (0.035 MPa).
All Series 7250 instruments up to 2,500 psi (17.2 MPa) reach setpoint in 20 seconds or less into a 15 cubic inch (245 cc) volume, with no overshoot, to allow high-speed pressure test and calibration.
The Series 7250 also provides two user-selectable control modes: active and passive mode. In active mode, the 7250 is continually maintaining the setpoint and can compensate for small leaks and pressure changes due to temperature. In passive mode, the user defines a control band and the 7250 will turn off the controller once it achieves the setpoint within the control band. In a leak-free and temperature- stable system, passive mode contributes no additional uncertainty, providing optimum performance.
The 7250 is easy to use and can automate your calibrations in several ways:
Computer interface: Every Series 7250 is provided with both an RS-232 and IEEE-488 interface, and Series 7250 syntax follows SCPI protocol for easy programming. A LabVIEW® driver is available as a free download. As a standard feature, software written for the previous generation Series 7215, 7010 and 6000 instruments is fully supported by the Series 7250. The Series 7250 can also be set to 510 emulation mode to use software originally written for the DPI 510. Firmware updates can also be performed over the RS-232 interface (updates can be downloaded from the web site). The 7250 series is compatible with COMPASS software.
The Series 7250 is versatile enough to handle almost any type of pneumatic pressure calibration.