Mineral-insulated sheath with a minimum bend radius of 19 mm (3/4 in) for flexibility and durability
Specifications
Resistance Nominal 100 Ω
Temperature coefficient 0.00385 Ω/Ω °C nominal
Temperature range -200 °C to 300 °C (transition and cable temperature: 0 °C to 150 °C)
Drift rate + 0.13 °C at 0 °C after 1000 hours at 300 °C
Sheath material 316 stainless steel
Leads Teflon™-insulated, nickel-plated stranded copper, 22 AWG
Termination 4-wire banana
Time constant Four seconds maximum for 63.2% response to step change in water moving at 3 fps
Bending radius 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip
Calibration Includes manufacturer’s NIST-traceable calibration and table with R vs. T values in 1 °C increments from -196 °C to 300 °C. Calendar – van Dusen coefficients included
Immersion At least 100 mm (4 in) recommended
Accuracy (includes calibration uncertainty and short-term stability)
± 0.050 °C at -196 °C
± 0.050 °C at 0 °C
± 0.051 °C at 200 °C
± 0.055 °C at 300 °C
Size 9 in L x 3/16 in diameter
Marking Each RTD is marked with the Fluke name and part number